Troubleshooting
Alarms and alerts
Message in display What it means / Alarm type How to make it stop
You are using the “Units of
insulin”-type bolus and the pump
has calculated that the “Current
Blood Glucose” you entered would
result in a blood glucose level
below your target, so no correction
bolus is suggested. Single alarm.
Due to “insulin on board” (from one or more
previous boluses which are still aecting your
blood glucose level or a low BG), the pump has
calculated no bolus is recommended at this
time. Recheck your blood glucose level and treat
a low as instructed on page 27 and according to
any instructions given to you by your healthcare
provider. Press Next to go to the bolus
conrmation screen, or press Back to return to
the current BG screen.
Delivery stops. e battery does
not have enough power for the
pump to work. Continuous alarm;
or, if power is too low, two beeps
repeated continuously.
You must change the battery immediately. If the
repeating two beep alarm occurs, power may be
so low that the display goes blank or the pump
continuously attempts to restart; always replace
the battery whenever this occurs.
is alert occurs only during
waking hours (6:00 to
10:00 ). A series of four
Attention alarms given every
four hours until battery is depleted.
Press OK. Change the battery as soon as
possible.
A custom alert is in use, and the
time period programmed has been
reached. Attention Alarm.
Press Snooze to quiet the alarm for ten minutes.
To cancel the alarm, press Cancel. Complete
whatever testing or other activity the alert was
used to remind you of.
While using the Basal Testing
feature, your blood glucose
became too high to safely continue
the testing. e basal test is halted.
Single Alarm.
Press OK. Treat the high blood glucose as
instructed on page 28 and according to any
instructions given to you by your healthcare
provider.
While using the Basal Testing
feature, your blood glucose became
too low to safely continue testing.
e basal test is halted. Single
Alarm.
Press OK. Treat the low blood glucose with fast-
acting carbohydrate as instructed on page 27,
and according to any instructions given to you
by your healthcare provider.
* Using the CoZmanager
®
2.0 soware, you can create your own custom message for this alarm. If the screen
has been customized, the message will be dierent than the one shown above.
Battery
depleted
Deliverystopped
Insertanew
battery
!
!
<Back Next>
Alert
BGisbelowtargetor
maydropbelow
target.Nobolus
suggested.
OK
Battery
Low
Changebattery
soon
!
!
<Snooze Cancel>
BG Alert X
*
OK>
BG Too
High
Basaltestended.
Followphysician’s
instructionstotreat
highBG.
!
OK>
BG Too
Low
Fastactingcarbs
recommended.
Followphysician’s
instructions.
!