85
Setup
You can set up to 24 correction factors, in half
hour increments. Factors will be in either mg/dl or
mmol/L, depending on which you use to measure
your blood glucose, and can be set as follows:
5 to 30 mg/dl in 1 mg/dl increments;
30 to 100 mg/dl in 5 mg/dl increments;
100 to 400 mg/dl in 10 mg/dl increments
– or –
0.2 to 1.5 mmol/L in increments of 0.05 mmol/L;
1.5 to 5 mmol/L in increments of 0.25 mmol/L;
5 to 23 mmol/L in increments of 0.5 mmol/L.
Specify your Duration of insulin action, or the
amount of time insulin remains active in your body
aer a bolus. is setting looks at previously deliv-
ered boluses and determines, based on the amount
of time you have programmed, if any insulin from
previous boluses is still actively aecting your blood
glucose level.
Once your duration of insulin action is correctly pro-
grammed and you use the correction bolus feature,
any insulin that is still active in your body from a
previous bolus will be taken into consideration. e
screen that tells you the bolus amount recommended
shows that the amount was reduced due to this
“insulin on board.”
Specify your Target Blood Glucose; you will work
with your healthcare provider to determine a target
blood glucose level specically for you. For some
people their target blood glucose varies throughout
the day and night, so the pump allows you to pro-
gram a single target for the whole day (starting at
midnight), or several targets, also starting at mid-
night, for specic times of the day or night.
You can set up to 24 target blood glucose settings:
mg/dl: 80 to 200 mg/dl in 5 mg/dl increments;
mmol/L: 4.4 to 11.1 mmol/L in 0.1 mmol/L
increments.
Basal patterns
Warning: You should receive specic training on
programming and using basal patterns. Please
carefully read all instructions and talk to your
healthcare provider.
You need to program a maximum basal rate and edit
your basal pattern(s).
In QuickStart™ you can edit the active basal pattern.
(Renaming the pattern(s) can only be done in
Setup > Delivery > Basal Pattern—see page 114.)
NOTE: If you personalize your pump in Quick-
Start™, then later enter other setup functions (such
as Setup > Delivery > Meal Bolus, for instance),
the settings shown in that menu may appear dif-
ferently, depending on the personalization done.
___________________________
•PressMenu.
•ChooseSetup then press
Select.
•WithQuickStarthighlighted
press Select.
•Ifyoudonotwanttoeditthe
time and date, with Nohigh-
lightedpressNext or
If you want to edit the time
and date choose Yes then
press Next.
<Back Next>
TimeandDate
Time:08:00AM
Date:05/12/06
Edit? No
k
<Home Select>
Basal Pattern
Load
History
Setup
##
m
<Back Select>
QuickStart
PumpSettings
Personalize
Delivery
SetupMenu
1
<Stop Menu>
Cozmo
293
u
##
Basal: Pattern 1
Rate:1.150u/hr