Using the Pump - More Features
95
Using the temporary rate
*This screen only appears if more than
1 temporary rate is available.
**If running a custom temporary rate,
the custom name is shown at the top of
this screen.
Cozmo
<Stop Menu>
Basal: Pattern 1
Rate: 1.150 u/hr
Tem porary
Rate
Duration: 01:30 hr
Rate: 1.10 u/hr
<Back Deliver
Select Temp orary Rate,
then press Select
Start at
Home screen
Cozmo
<Stop Menu>
Basal: Pattern 1
Rate: 0.750 u/hr
Tem p Rate Active
Press
Menu
Select type,
then press Select *
Tem porary Rate
Menu
Standard
Exercise
Temp Rate 3
<Back Select>
Tem porary
Rate
Duration: 01:30 hr
Rate: 1.10 u/hr
<Back Deliver
Select Duration (00:00 to 72:00
hours), then press Next
Verify Duration and Rate, then press Deliver
**
Select Rate (Percent: 10% to 250% or %
which equals basal rate maximum;
Units/Hour: 0.00 to basal rate maximum),
then press Next
Tem porary
Rate
Duration: 00:30 hr
Rate: 100 %
<Back Next>
Tem porary
Rate
Duration: 00:30 hr
Rate: 0.00 u/hr
<Back Next>
Tem porary
Rate
Duration: 01:30 hr
Rate: 50 %
<Back Next>
Tem porary
Rate
Duration: 01:30 hr
Rate: 1.10 u/hr
<Back Next>
Meal Bolus
Correction Bolus
Tem porary Rate
Basal Pattern
<Home Select>