Troubleshooting
Alarms
Message in display What it means / Alarm How to make it stop
Battery
depleted
Delivery stopped
Insert a new battery
!
Blockage
detected
No delivery. Check
tubing or site for
blockage
!
Cartridge
empty
Insert filled cartridge
to continue delivery
!
Delivery stops. The battery does not
have enough power for the pump to
work. Continuous alarm; or, if
power is too low, 2 beeps repeated
continuously.
Delivery stops. There is something
preventing insulin from being deliv-
ered. Continuous alarm.
Delivery stops. The cartridge is
empty. Continuous alarm.
You must change the battery immediately. If the
repeating 2 beep alarm occurs, power may be so
low that the display goes blank or the pump
continually attempts to restart; always replace
the battery whenever this occurs.
Press OK to silence alarm and stop delivery.
Alarm will continue to occur until blockage is
cleared. Disconnect your infusion set from the
access device until the blockage is cleared. Check
tubing, making sure it is not kinked; make sure
tubing is not trapped in clothing or pouch.
Check access site, making sure access device is
properly inserted. If still blocked, change access
device, infusion set, and access site. Test your BG
level. Restart delivery.
Press OK. You must load a new cartridge now.