PROOF (LR #3557), 2002-11-25 «5118-51A Cozmo User Manual»
Alerts - continued
Message in display What it means / Alarm How to make it stop
Alert
BG is below target or
may drop below
target. No bolus
suggested.
!
No bolus
allowed
Bolus in progress.
!
No bolus
allowed
Cartridge volume is
very low.
!
No bolus
allowed
Pump is stopped.
!
Missed meal
bolus
No bolus delivered
between xx:xx and xx:xx.
Q
The pump has calculated that the
“Current Blood Glucose” you entered
would result in a BG level below
your target. No correction bolus will
be suggested. Attention alarm.
A missed meal bolus alert was pro-
grammed, and no bolus was deliv-
ered during the selected time inter-
val. Attention alarm.
You attempted to deliver a bolus
while a bolus was delivering. Single
alarm.
You attempted to deliver a bolus, but
the cartridge volume is too low.
Single alarm.
You attempted to deliver a bolus
while the pump was stopped. Single
alarm.
Press OK. Due to “insulin on board” (from 1 or
more previous boluses which are still affecting
your BG level), the pump has calculated a 0
(zero) bolus at this time. Recheck your BG level
and treat a low BG as instructed on page 67 and
according to any instructions given to you by
your healthcare provider.
Press OK. If you have eaten and forgotten your
bolus, check your BG level immediately. Deliver
a bolus, if required.
Press OK. Only 1 bolus can be delivered at a
time.
Press OK. Load a new cartridge, then deliver the
bolus.
Press OK. Restart the pump, then deliver the
bolus.