Deltec Cozmo Bedienungsanleitung Seite 123

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 137
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 122
Technical Information
111
Basal Pattern
User can edit (4 patterns, each with a possible 48 time/rate
segments) and start using basal patterns
Delivery Specifications:
Time: Midnight to 11:30PM in 00:30 increments or
Midnight to 23:30 in 00:30 increments
Default: --:--
Rate: 0.00 u/hr (units/hour) or 0.05 u/hr to basal rate
maximum
Default: 0.00 u/hr
Setup in Pump Settings / Personalize / Delivery / Basal
Pattern menu
Number of patterns: 1 to 4 patterns may be displayed or
not displayed
Default: 1 pattern displayed
Pattern names: Pattern “X” (1, 2, 3, 4), Exercise day,
Weekday, Travel, Weekend, Monthly
Default: Pattern “X” (1, 2, 3, 4)
Basal rate maximum: 0.05 u/hr to 35.00 u/hr in 0.05 u/hr
increments
Default: 2.00 u/hr
Load Cartridge
User loads cartridge, fills tubing, fills cannula, and selects
site change alarm (if displayed)
Fill tubing: Fills in 1.0 unit increments until Stop is
pressed (automatically stops at 30 units).
Fill cannula: 0.10 units to 1.0 units in 0.1 unit incre-
ments
Default: 0.0 units initially, then last programmed
amount
Site change alert (if displayed - see Alert Specifications):
1 to 5 days in 1 day increments
Default: 3 days
12:30AM (00:30) to 11:30PM (23:30) in 00:30 minute
increments
Default: 8:00 AM
History
User selects displayed report for viewing
View:
Memory:
Reports: last 90 days, arranged from most recent to oldest
Events: last 2000 events, arranged from most recent to
oldest
Seitenansicht 122