PROOF (LR #3557), 2002-11-25 «5118-51A Cozmo User Manual»
Alerts - continued
Message in display What it means / Alarm How to make it stop
Check blood
glucose
It has been x:xx hours
since your last bolus
or cartridge change
Q
Cartridge
Volume Low
Change cartridge
soon
!
Delivery
is stopped
!
Delivery Limit
No bolus
xxx units already delivered
in past 60 min. Bolus
would exceed limit.
!
Extended delivery in
progress - No
combo or extended
bolus allowed
!
The amount of insulin left in the car-
tridge is low. Alert repeats at intervals
until Cartridge Very Low alarm oc-
curs. Attention alarm.
It is time to test your BG level to
make sure it is within your target
range. Attention alert.
The pump is stopped, and no insulin
is being delivered. Stopped alarm.
The bolus you attempted was not
delivered as it would exceed the
delivery limit. Attention alert.
You attempted to enter the extended
or combination bolus menu when an
extended bolus is being delivered.
Single alarm.
Press OK. Replace the cartridge as soon as
practical.
Press OK. Perform BG testing. If it is within your
target range, do nothing. See the high and low
blood glucose sections in this chapter if your BG
level is outside your target range.
Press OK. Alert screen and alarm (beep or
vibration) occur every 5 minutes until Resume is
pressed to restart insulin delivery.
Press OK. You must wait to deliver the bolus, or
contact your healthcare provider to discuss your
delivery limit.
Press OK. Only 1 extended bolus can be deliv-
ered at a time. If you want to stop the current
extended bolus, from the home screen press
Stop, then choose the type of delivery you want
to stop.